宁霄, 金绍明, 孙雯雯, 曹进, 裴宇盛. 高效液相色谱-串联质谱法同步测定饮用水中18种常见消毒副产物的方法研究[J]. 上海预防医学, 2023, 35(7): 711-718. DOI: 10.19428/j.cnki.sjpm.2023.22731
引用本文: 宁霄, 金绍明, 孙雯雯, 曹进, 裴宇盛. 高效液相色谱-串联质谱法同步测定饮用水中18种常见消毒副产物的方法研究[J]. 上海预防医学, 2023, 35(7): 711-718. DOI: 10.19428/j.cnki.sjpm.2023.22731
NING Xiao, JIN Shaoming, SUN Wenwen, CAO Jin, PEI Yusheng. Simultaneous determination of 18 common disinfection by-product residues in drinking water by high performance liquid chromatography-tandem mass spectrometry[J]. Shanghai Journal of Preventive Medicine, 2023, 35(7): 711-718. DOI: 10.19428/j.cnki.sjpm.2023.22731
Citation: NING Xiao, JIN Shaoming, SUN Wenwen, CAO Jin, PEI Yusheng. Simultaneous determination of 18 common disinfection by-product residues in drinking water by high performance liquid chromatography-tandem mass spectrometry[J]. Shanghai Journal of Preventive Medicine, 2023, 35(7): 711-718. DOI: 10.19428/j.cnki.sjpm.2023.22731

高效液相色谱-串联质谱法同步测定饮用水中18种常见消毒副产物的方法研究

Simultaneous determination of 18 common disinfection by-product residues in drinking water by high performance liquid chromatography-tandem mass spectrometry

  • 摘要:
    目的 基于超高效液相色谱-串联质谱(UHPLC⁃MS/MS)建立饮用水中18种常见卤氧化物和卤代乙酸消毒副产物的快速检测方法。
    方法 水样经0.22 μm亲水滤膜过滤后,采用PFP(2.1 mm× 100 mm,2.7 μm)五氟苯基色谱柱分离,0.1%乙酸和乙腈为流动相梯度洗脱,在阴离子电喷雾模式下电离,采用多反应监测(MRM)模式检测,外标法定量。
    结果 4种卤氧化物和14种卤乙酸均在相应线性范围内,相关系数均>0.999。每种分析物1、10和20倍定量限浓度水平的平均加标回收率为91.6%~101.8%,相对标准偏差(RSD)为1.2%~6.4%。方法检出限和定量限分别为0.020~2 μg·L-1和0.050~5 μg·L-1
    结论 此方法快速简便、灵敏度高、准确有效,可用于饮用水中18种常见消毒副产物的快速检测。

     

    Abstract:
    Objective A method was developed for the rapid determination of 18 common disinfection by-products including halogenated oxides and haloacetic acid (HAAs) in drinking water by high performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS).
    Method The water sample was filtered by 0.22 μm hydrophilic membrane then the analytes were separated on a PFP (2.1 mm× 100 mm, 2.7 μm) pentafluorophenyl column with 0.1% acetic acid and acetonitrile as mobile phase gradient elution. Ionization in anionic electrospray mode was detected by multi-reaction monitoring (MRM) mode. The external standard method was used for quantitation.
    Results The correlation coefficients of 18 disinfection by-products were above 0.999 in the corresponding linear range. The average spiked recoveries of 1, 10 and 20 times of LOQ of each analyte were 91.6%‒101.8%, and the relative standard deviation (RSD) was 1.2%‒6.4%. The LOD and LOQ were 0.020‒2 μg·L-1 and 0.050‒5 μg·L-1, respectively.
    Conclusion This method is simple, sensitive and accurate, and could be used for the routine analysis of 18 common disinfection by-products in drinking water.

     

/